Description
Full PTFE housing top.
Maximum chemical resistance, for use in highly aggressive mediums.
Filter fineness of 2 μm or 25 μm with sintered PTFE elements.
Even if the particular contaminant has already been removed at the extraction point through effective particle filtration inside the gas sampling probe, in long or branched sample gas line there is a risk of secondary contamination. Additional filters are therefore often installed at the point where the sample gas enters the analysis system and also before delicate system components. The filter housings must be made of corrosion-resistant, non-absorbent materials, easy to install and easy to maintain. They should further be compatible with various filter elements. The AGF-T- 30 series has a PTFE and glass housing. The filter head has an additional connection for installing a moisture detector. A variety of materials is available based on the application.
Full PTFE housing top.
Maximum chemical resistance, for use in highly aggressive mediums.
Filter fineness of 2 μm or 25 μm with sintered PTFE elements.
No absorption of trace gasses.
Quick response time in systems due to low dead volume.
Variable wall mount.
Material – Filter head | PTFE |
Material – Filter cover | Glass |
Material – Gasket | Viton, PTFE shrouded |
Operating temperature max. | 100 °C |
Operating pressure max. | 4 bar |
Datasheet |
|